VITA Update Messenger
| 24.02.2025

Chère Madame, cher Monsieur,
The new software updates are now available for:
Software-Update 25011
The new software updates are now available for:
Software-Update 25011
- VITA V60 i-Line PLUS
- VITA vPad excellence as of serial number 2520172050
- VITA vPad comfort as of serial number 3020180051
- VITA VACUMAT 6100 M
- VITA SMART.FIRE ADVANCED
Software Update Version 250117

VITA V60 i-Line PLUS
Software Update Version 250203

VITA vPad excellence as of serial number 2820170050

VITA vPad comfort as of serial number 3020180051

VITA VACUMAT 6100 M

VITA SMART.FIRE ADVANCED
Information on the content
Software-Update 250117
Software-Update 250203
VITA vPad comfort / excellence
VITA VACUMAT 6100 M / VITA SMART.FIRE ADVANCED
- Turkish has been integrated into the user interface as an additional language.
- The control unit has been optimized to allow network fluctuations to be handled better.
- The vacuum pump will now be reliably activated, even if the vacuum is only required during the holding phase.
- Anticon will now be correctly activated, even if the device switches to standby mode immediately after switching on.
- Improved stability during a data export via USB drive.
Software-Update 250203
VITA vPad comfort / excellence
- The available storage space for process data is now checked every 24 hours. This prevents problems that could occur if the device has never been switched off.
- Improved stability during a data export via USB drive.
- “YZ MULTI TRANSLUCENT Universal” and “YZ MULTI TRANSLUCENT Speed” sintering programs added.
- SUPRINITY PC glaze firing added.
- Processing instructions updated.
VITA VACUMAT 6100 M / VITA SMART.FIRE ADVANCED
- Anticon will now be correctly activated, even if the device switches to standby mode immediately after switching on.
- The vacuum pump will now be reliably activated, even if the vacuum is only required during the holding phase.
- The control of TRIAC components has been optimized to allow network fluctuations to be handled better.
- Faults in the control of the Multipump have been corrected.